9140 Series
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 9142 | 9143 | 9144 | |
| Temperature Range (at 23°C ambient) | –13 to 302°F –25 to 150°C |
91 to 662°F 33 to 350°C |
122 to 1220°F 50 to 660°C |
| Display Accuracy | ±0.2°C Full Range | ±0.2°C Full Range | ±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C |
| Stability | ±0.01°C Full Range | ±0.02°C at 33°C ±0.02°C at 200°C ±0.03°C at 350°C |
±0.03°C at 50°C ±0.05°C at 420°C ±0.05°C at 660°C |
| Axial Uniformity at 40 mm (1.6 in) | ±0.05°C Full Range | ±0.04°C at 33°C ±0.1°C at 200°C ±0.2°C at 350°C |
±0.05°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C |
| Radial Uniformity | ±0.01°C Full Range | ±0.01°C at 33°C ±0.015°C at 200°C ±0.02°C at 350°C |
±0.02°C at 50°C ±0.05°C at 420°C ±0.10°C at 660°C |
| Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) | ±0.006°C Full Range | ±0.015°C Full Range | ±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C |
| Hysteresis | 0.025°C | 0.03°C | 0.1°C |
| Operating Conditions | 0 to 50°C, 0% to 90% RH (non-condensing) | ||
| Environmental Conditions (for all specifications except temperature range) | 55 to 91ºF 13 to 33°C |
||
| Immersion (Well) Depth | 5.9 in. / 150mm | ||
| Insert OD | 1.18in / 30mm | 1.00in / 25.3mm | 0.96in / 24.4mm |
| Heating Time | 16 min: 23 to 140°C 23 min: 23 to 150°C 25 min: –25 to 150°C |
5 min: 33 to 350°C | 15 min: 50 to 660°C |
| Cooling Time | 15 min: 23 to –25°C 25 min: 150 to –23°C |
32 min: 350 to 33°C 14 min: 350 to 100°C |
35 min: 660 to 50°C 25 min: 660 to 100°C |
| Resolution | 0.01 ° | ||
| Display | LCD, °C or °F user-selectable | ||
| Size (H x W x D) | 11.4 (H) x 7.3 (W) x 11.6 (D) in. / 290 (H) x 185 (W) x 295 (D) mm | ||
| Weight | 18 lb / 8.16 kg | 16 lb / 7.3 kg | 17 lb / 7.7 kg |
| Power Requirements | 100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W |
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W |
|
| Computer Interface | RS-232 and 9930 Interface-it control software included | ||
-P Specifications:
| Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe)† | ±0.013°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C |
| Reference Resistance Range | 0 ohms to 400 ohms |
| Reference Resistance Accuracy‡ | 0 ohms to 42 ohms: ±0.0025
ohms 42 ohms to 400 ohms: ±60 ppm of reading |
| Reference Characterizations | ITS-90, CVD, IEC-751, Resistance |
| Reference Measurement Capability | 4-wire |
| Reference Probe Connection | 6-pin Din with Infocon Technology |
| Built-in RTD Thermometer Readout Accuracy | NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C |
| RTD Resistance Range | 0 ohms to 400 ohms |
| RTD Resistance Accuracy‡ | 0 ohms to 25 ohms: ±0.002
ohms 25 ohms to 400 ohms: ±80 ppm of reading |
| RTD Characterizations | PT-100 (385),(JIS),(3926), NI-120, Resistance |
| RTD Measurement Capability | 4-wire RTD (2-,3-wire RTD w\ Jumpers only) |
| RTD Connection | 4 terminal input |
| Built-in TC Thermometer Readout Accuracy | Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±0.6°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C |
| TC Millivolt Range | –10 mV to 75 mV |
| Voltage Accuracy | 0.025% of reading + 0.01 mV |
| Internal Cold Junction Compensation Accuracy | ± 0.35°C (ambient of 13 to 33°C) |
| TC Connection | Small connectors |
| Built-in mA Readout Accuracy | 0.02% of reading + 0.002 mA |
| mA Range | Cal 4-22 mA, Spec 4-24 mA |
| mA Connection | 2 terminal input |
| Loop Power Function | 24 V dc loop power |
| Built-in Electronics Temperature Coefficient (0 to 13°C, 33 to 50°C) | ±0.005 % of range per °C |
| † The temperature range
may be limited by the reference probe connected
to the readout. The Built-In Reference
Thermometer Readout Accuracy does not include
the sensor probe accuracy. It does not include
the probe uncertainty or probe characterization
errors. ‡ Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. |
|
The 9140 series display panel.
Choose any of six standard inserts.
| 9142 | |||
| 9142-X* | Field Metrology Well 9142 and one insert, -25 to 150ºC |
||
| 9142-X*-P | Field Metrology Well 9142 and one insert, -25 to 150ºC, with Process Electronics |
||
| 9143 | |||
| 9143-X* | Field Metrology Well 9143 and one insert, 33 to 350ºC |
||
| 9143-X*-P | Field Metrology Well 9142 and one insert, 33 to 350ºC, with Process Electronics |
||
| 9144 | |||
| 9144-X* | Field Metrology Well 9144 and one insert, 50 to 660ºC, |
||
| 9144-X*-P | Field Metrology Well 9144 and one insert, 50 to 660ºC, with Process Electronics |
||
| Options | |||
| 9142-INSX* | Optional Inserts for 9142. |
||
| 9143-INSX* | Optional Inserts for 9143. |
||
| 9144-INSX* | Optional Inserts for 9144. |
||
| To order more than one insert for a particular model, return to this page after adding the first one. | |||
| 9142-CASE | Rugged Carrying Case | ||
| *The X designation in the model number represents your choice of insert A, B, C, D, E, or F. | |||