Hart Scientific
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| PRT / RTD | Thermistor | Thermocouple | |
| Inputs: |
2 channels PRT/thermistor and 2 channels TC, or 4 channels PRT/thermistor, or 4 channels TC, specify when ordering;
PRT/thermistor channels accept 2, 3, or 4 wires; TC inputs accept B, E, J, K, N, R, S, T, and Au-Pt TC types. (Support for C and U type thermocouples is available. Download the application note Using Hart Readouts with Tungsten-Rhenium and other Thermocouples from www.hartscientific.com.) |
||
| Temperature Range: | 189°C to 960 °C | 50°C to 150°C | 270°C to 1800°C |
| Measurement Range: |
0 to 400 Ω
|
0 to 500 KΩ
|
10 to 100 mV
|
| Characterizations: |
ITS-90, IEC-751 (DIN “385”), Callendar-Van Dusen
|
Steinhart-Hart, YSI-400
|
NIST Monograph 175, 3-point deviation function applied to NIST 175, 6th-order polynomial
|
| Temperature Accuracy (meter only): | ±0.004°C at 100°C ±0.006°C at 0°C ±0.009°C at 100°C ±0.012°C at 200°C ±0.018°C at 400°C ±0.024°C at 600°C |
±0.0025°C at 0°C
±0.0025°C at 25°C ±0.004°C at 50°C ±0.010°C at 75°C ±0.025°C at 100°C |
Ext. RJC / Int. RJC
B at 1000°C ±0.6°C / ±0.6°C; E at 600°C ±0.07°C / ±0.25°C; J at 600°C ±0.1°C / ±0.35°C; K at 600°C ±0.15°C / ±0.4°C; N at 600°C ±0.15°C / ±0.3°C; R at 1000°C ±0.4°C / ±0.5°C; S at 1000°C ±0.5°C / ±0.6°C; T at 200°C ±0.1°C / ±0.3°C |
| Temperature Resolution: |
0.001°
|
0.0001°
|
0.01 to 0.001°
|
| Resistance/Voltage Accuracy: |
0Ω to 20Ω: ±0.0005Ω
20Ω to 400Ω: ±25 ppm of rdg. |
0Ω to 5 KΩ: ±0.5Ω
5 KΩ to 200 KΩ: ±100 ppm of rdg. 200 KΩ to 500 KΩ ±300 ppm of rdg. |
10 to 50 mV: ±0.005 mV
50 to 100 mV: ±100 ppm of rdg. (Internal RJC: ±0.25 °C) |
| Environmental Range: | 16°C to 30°C | ||
| Measurement Interval: |
0.1 second to 1 hour; inputs may be read sequentially or simultaneously at 1 second or greater interval
|
||
| Excitation Current: |
1 µA, reversing
|
2 and 10 µA, automatically selected
|
n/a
|
| Display: |
33 x 127 mm (1.3 x 5 in) backlit LCD graphical display
|
||
| Display Units: |
°C, °F, K, Ω, KΩ, mV
|
||
| Data Logging: |
Up to 8,000 time- and date-stamped measurements can be logged
|
||
| Logging Intervals: |
0.1, 0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes
|
||
| Averaging: |
Moving average of most recent 2 to 10 readings, user selectable
|
||
| Probe Connection: |
Patented DWF Connectors accept mini spade lug, bare-wire, or mini banana plug terminations
|
Universal receptacle accepts miniature and standard TC connectors
|
|
| Communications: |
RS-232 included, IEEE-488 (GPIB) optional
|
||
| AC Power: |
100240 VAC, 50-60 Hz, 0.4 A
|
||
| DC Power: |
1216 VDC, 0.5 A (battery charges during operation from 14.5 to 16V DC, 1.0A)
|
||
| Battery: |
NiMH, 8 hours of operation typical without backlight, 3 hours to charge, 500 cycles
|
||
| Size (WxHxD): | 102 x 191 x 208 mm (4.0 x 7.5 x 8.2 in) | ||
| Weight: | 2 kg (4.5 lb.) | ||
| NIST- Traceable Calibration: |
Accredited NIST-Traceable resistance calibration and NIST-Traceable voltage calibration provided | ||
1529 Chub-E4 Thermometer back panel
1529 Chub-E4 Thermometer,
2 TC and 2 PRT/Thermistor inputs
1529-R Chub-E4 Thermometer,
4 PRT inputs
1529-T Chub-E4 Thermometer,
4 TC inputs
The 9322 rugged carrying case holds your 1529 and four probes up to 12" long
To Order:
| 1529 | Chub-E4 Thermometer, 2 TC and 2 PRT/Thermistor inputs |
|
|
| 1529-R | Chub-E4 Thermometer, 4 PRT/Thermistor inputs |
|
|
| 1529-T | Chub-E4 Thermometer, 4 TC inputs |
|
|
| 2506-1529 | IEEE Option | ||
| 9322 | Rugged Carrying Case, holds 1529 and four probes up to 12" long | ||
| 2513-1529 | Rack-Mount Kit | ||
| 9935-S | LogWare II, Multi Channel, Single User | ||
| 9935-M | LogWare II, Multi Channel, Multi User | ||
| 9320A | External Battery Pack |