Hart Scientific 9150 Thermocouple Furnace

Model #9150-X

Low-cost thermocouple furnace
NIST-traceable calibration included
RS-232 port standard
Range to 2,192°F (1200°C)

From: To Order

Bookmark and Share

Click and hover to zoom. Click again to view larger image.

You told us you weren't satisfied with the competition's furnaces for checking industrial thermocouples. You said you wanted something new and more convenient to use—and you wanted it at a lower price than any other furnace available. Well, we've got what you asked for, and it's the Model 9150 Thermocouple Furnace from Hart Scientific.

With a stability of ±0.5 °C, it has a temperature range to 1200 °C and a display accuracy of ±5 °C across its entire range.

With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter. The 9150 works with 115 or 230 VAC power.

The 9150 Thermocouple Furnace uses Hart's own microprocessor-based controller for great stability and set-point accuracy. It has a removable well insert for versatility. It has rapid cool-down and heat-up times. And it comes with an RS-232 port for connection to a PC.

You can now afford to check your thermocouples with this excellent cost-effective instrument. Why pay more for features you don't need and can't use? Each unit is factory-calibrated and comes with test data and a calibration traceable to NIST.

Range 302 to 2192°F (150 to 1200°C)
Accuracy ±5°C
Resolution 0.1° to 999.9°, 1° above 1000°
Power 115VAC (±10%), 10.5A or 230VAC (±10%), 5.2A, switchable, 50/60Hz, 1200W
Stability ±0.5°C
Well Diameter 1.25" (32 mm)
Well Depth 5.5" (140 mm); (4" [101 mm] removable insert plus 1.5" [38 mm] insulator)
Heating Time 35 minutes to 1200°C
Cooling Time 140 minutes with block
Well-to-Well Uniformity ±0.5°C to ±1.0°C (Insert "C" at 1200°C)
Stabilization ±0.5 °C
Dimensions 12.4" x 8.2" x 12.4" (315 x 208 x 315 mm)
Weight 28 lb. (13 kg)
Calibration Data at 150°C, 300°C, 450°C, 600°C, 800°C, 1000°C, and 1200°C

 


9150 Insert Options
A: 1/16", 1/8", 3/16", 3/8", 1/2"
B: 3/16" x 2, 1/4" x 2, 3/8" x 2
C: 1/4" x 6
D: 3mm x 2, 4mm x 2, 6mm x 2

Hart Scientific 9150 Furnace
9150-X Thermocouple Furnace (specify X, X = A, B, C,or D included insert)
3150-2 Insert A  
3150-3 Insert B  
3150-4 Insert C  
3150-6 Insert D  
9315 Rugged Carrying Case  

Our Company