High-Temp Secondary Standard PRT Temperature Sensors

Model #5626 / 5628

Range to 660°C
Meets all ITS-90 requirements for resistance ratios
RTPW drift < 20mK after 500 hours at 660°C
Strain-free sensor design by Hart's SPRT engineers

From: To Order

Bookmark and Share

Click and hover to zoom. Click again to view larger image.

Hart's high-temp secondary standards fill the gap between affordable, but temperature-limited secondary PRTs and more expensive, highly accurate SPRTs.

If you're using block calibrators, furnaces, or temperature points above normal PRT temperatures (420 °C), then these two PRTs are for you. The 5626 is nominally 100W and the 5628 is nominally 25.5W. Both instruments have a temperature range of –200 °C to 661 °C. They make great working or check standards for calibration work up to the aluminum point.

Using a regular PRT at temperatures above 500 °C exposes the platinum to contamination. If the PRT is used as a reference or calibration standard, contamination is a major problem. SPRTs, which are more expensive and delicate, can handle the higher temperatures, but with greater risk to the instrument due to shock, contamination, or mishandling. The 5626 and 5628 are designed to reduce the contamination risk through the use of internal protection while not impairing performance.

In addition to the right measurement performance and durability, a PRT for secondary applications should be priced affordably. Hart's new PRTs are inexpensive and come with an accredited calibration. The calibration comes complete with ITS-90 constants and a resistance-versus-temperature table.

Check the temperature range, check the stability, check the price! Who else gives you this much quality, performance, and value for your money? No one!

Range –200 to 661°C
Handle Temp. 0 to 80°C
RTPW 5626: 100W (±1W)
5628: 25.5W (±0.5W)
W(Ga) ³ 1.11807
Calibration Uncertainty (k=2) 0.006°C at –200°C
±0.004°C at 0°C
±0.009°C at 420°C
±0.014°C at 661°C
Stability 5626: ±0.003°C
5628: ±0.002°C
Long-Term Drift 5626: < 0.03°C/500 hours at 661°C
5628: < 0.02°C/500 hours at 661°C
Immersion At least 5" (12.7 cm) recommended
Sheath Inconel™ 600
Lead Wires 4-wire Super-Flex PVC, 22 AGW
Termination Gold-plated spade lugs, or specify
Dimensions* 0.25" x 12", 15", or 20" (6.35 mm dia. x 305 mm, 381 mm, or 508 mm) standard, custom lengths available
Certificate*** Accreditated calibration from Fluke Hart Scientific

To Order:

Hart Scientific Secondary PRT Temperature Sensors
5626-12-X  High-temp PRT, 100W, 305 mm (12 in) 
5626-15-X  High-temp PRT, 100W, 381 mm (15 in) 
5626-20-X  High-temp PRT, 100W, 508 mm (20 in) 
5628-12-X  High-temp PRT, 25.5W, 305 mm (12 in) 
5628-15-X  High-temp PRT, 25.5W, 381 mm (15 in) 
5628-20-X  High-temp PRT, 25.5W, 508 mm (20 in) 
2609  Spare Case   
Appropriate case included with purchase of 5626 or 5628 PRT.

X = termination. Specify "L" (mini spade lugs), "D" (5-pin DIN for Tweener Thermometers), "P" (Smart Lemo), or "S" (spade lugs).