Practical Instrument Electronics 520 Thermocouple Simulator
- Perfect for checking the accuracy of any thermocouple meter.
- Easy-to-set digital knob dials-in any temperature quickly.
- Low price model dedicated to one TC type of your choice.
- Typical accuracy is ±0.25°C.
|
If you use thermocouple meters, you need a way to check their accuracy. This brand new calibrator offers a great value for the price and you wont find anything easier-to-use. Quickly dial-in any temperature in 1° increments using the digital dial. Press the knob as you turn and you get even faster dialing. Easily beats the complexity and slowness of other digital calibrators. Youll get to your precise setting faster with no diddling around. Theres no complicated programming or tedious manual-reading. The EZ-Check function stores three temperatures of your choice for instant recall. Use these for fast checking of zero and span or any critical temperatures that are used repeatedly. The 520 is offered in eight popular thermocouple types. Choose the one you need when ordering. Each unit is fitted with a common mini-jack. A coiled connecting cable is available separately or use your own cables. A factory certificate of compliance is shipped with each unit. An optional traceable certificate may be ordered with actual test data.
| Specifications |
|
| Range: |
-148°F to full scale for the selected thermocouple type |
| Resolution: |
1°F or 1°C |
| Accuracy: |
|
| Millivolt Uncertainty: |
±(0.015% of mV Setting +0.009mV) |
| Output Noise: |
±5microVpp from 0.1Hz to 10Hz |
| Output Impedence: |
0.2ohms |
| Cold Junction Uncertainty: |
±0.5°C (0.9°F) |
| Cold Junct. Sensor Coefficient: |
±0.05°C/° in ambient temperature (°C or °F) |
| General Temp Accuracy: |
±(0.015% of mV setting + 0.009mV) ±0.5°C (0.9°F) |
| Operating Temp Range: |
-25 to 60°C (-10 to 140°F) |
| Storage Temp Range: |
-25 to 60°C (-10 to 140°F) |
| Battery: |
9V, 45 hours continuous use |
To Order:
| PIE-520-MJ |
PIE 520 Type J Thermocouple Simulator |
|
|
| PIE-520-MK |
PIE 520 Type K Thermocouple Simulator |
|
|
| PIE-520-ME |
PIE 520 Type E Thermocouple Simulator |
|
|
| PIE-520-MT |
PIE 520 Type T Thermocouple Simulator |
|
|
| PIE-520-MR |
PIE 520 Type R Thermocouple Simulator |
|
|
| PIE-520-MS |
PIE 520 Type S Thermocouple Simulator |
|
|
| PIE-520-MN |
PIE 520 Type N Thermocouple Simulator |
|
|
| PIE-520-MmV |
PIE 520 Type mV Thermocouple Simulator |
|
|
Coiled Lead
Use these to connect the simulator to a thermometer. Match the lead type to your meter. Length is 39" (1 meter) from plug to plug. |
| XCK-1M-MM |
Coiled Lead, 39" (1 meter), Type K, ANSI color Yellow |
|
|
| XCT-1M-MM |
Coiled Lead, 39" (1 meter), Type T, ANSI color Blue |
|
|
| XCJ-1M-MM |
Coiled Lead, 39" (1 meter), Type J, ANSI color Black |
|
|
|
|