Harts Micro-Baths are the worlds first completely portable baths built specifically for temperature calibration work. They cover a temperature range from -30°C to 200°C (-22°F to 392°F) to cover most industrial temperature applications.
Micro-Baths can be used anywhere for any type of sensor. The Model 6102 weighs less than 10 pounds, with the fluid. Its lighter and smaller than most dry-wells, has a spill-proof lid, and is easier to carry than your lunch. You can take it where you need to go without carts or excessive effort. Micro-Baths can even be transported with the fluid in them.
Wherever you go with your Micro-Bath, you can count on its performance. Each model is stable to ±0.03°C or better, depending on the fluid you use. Uniformity is ±0.02°C or better for low uncertainties using a reference thermometer. Display accuracy has been improved to ±0.25°C for quick calibrations without a reference thermometer. In short, you get the stability and precision of a liquid bath in a dry-well-sized package. Dont be fooled by competitors who place a can of oil inside a dry-well and call it a bath. Hart Micro-Baths are maximized for true fluid-bath performance.
With a 1.9-inch-diameter, 5.5-inch-deep tank, a Micro-Bath can calibrate any type of sensor including short, square, or odd-shaped sensors. The problems of fit and immersion are virtually eliminated by using a fluid medium rather than a dry-block calibrator. Micro-Baths are perfect for liquid-in-glass and bimetal thermometers.
The 6102 has a temperature range from 35°C to 200°C, the 7102 covers 5°C to 125°C, and the 7103 extends from 30°C to 125°C. Stability, uniformity, and accuracy specifications cover the entire range for each bath, not just the best temperature.
All Micro-Baths have RS-232 ports, come with our Interface-it software, and can be used with Harts MET/TEMP II software. Also included are contacts to calibrate a thermal switch, eight set-point memory storage, ramp-rate adjust, and over-temperature safety cutout.
You may have noticed we havent touted our CFC-free refrigeration. Yes, cold Micro-Baths are CFC-free, and also compressor-free. Thats rightno heavy, noisy compressor to lug around. We achieve our temperature range and stability with only one moving part. This means more durability and less weight.
We use Hart Scientific equipment in our lab here at ThermoWorks so we can tell you anything you need to know about how they work. Were not your average reseller; we use the stuff, understand how it works, can recalibrate most of it, and offer added technical tips to make your experience with the equipment even better.
|
Specifications
|
7102
|
7103
|
6102
|
|
Range:
|
5°C to 125°C
(23°F to 257°F)
|
30°C to 125°C
(22°F to 257°F)
|
35°C to 200°C
(95°F to 392°F)
|
|
Accuracy:
|
±0.25°C
|
|
Stability:
|
±0.015°C at -5°C
±0.03°C at 121°C
|
±0.03°C at -25°C
±0.05°C at 125°C
|
±0.02°C at 100°C
±0.03°C at 200°C
|
|
Uniformity:
|
±0.02°C
|
|
Resolution:
|
0.01°C/F
|
|
Operating Temp:
|
5°C to 45°C
|
|
Heating Time:
|
25°C to 100°C: 30 minutes
|
25°C to 100°C: 35 minutes
|
25°C to 200°C: 40 minutes
|
|
Cooling Time:
|
25°C to 0°C: 30 minutes
|
25°C to -20°C: 45 minutes
|
200°C to 100°C: 35 minutes
|
|
Well Size:
|
1.9" dia. x 5.5" deep (48 x 139 mm) |
|
Size:
|
7.2" W x 12" H x 9.5" D
(18 x 31 x 24 cm)
|
9" W x 13.2" H x 10.5" D
(23 x 34 x 26 cm)
|
5.5" W x 10.38" H x 8" D
(14 x 26 x 20 cm)
|
|
Weight:
|
15 lb. (6.8 kg) with fluid
|
22 lb. (9.8 kg) with fluid
|
10 lb. (4.5 kg) with fluid
|
|
Power:
|
115 VAC (±10%), 1.8 A or 230 VAC (±10%), 0.9 A, switchable, 50/60 Hz, 200W
|
94234 VAC (±10%), 50/60 Hz, 400W
|
115 VAC (±10%), 2.3 A or 230 VAC (±10%), 1.1 A, switchable, 50/60 Hz, 270W
|
|
NIST-
Traceable Calibration:
|
Data at -5°C, 25°C, 55°, 90°C, 121°C
|
Data at -25°C, 0°C, 25°C, 50°C, 75°C, 100°C, 125°C
|
Data at 50°C, 100°C, 150°C, 200°C
|
To Order:
| 6102 |
| 6102 |
Micro-Bath, 35°C to 200°C (95°F to 392°F) |
|
|
| 2082-M |
Spare Test Lid, 6102 |
|
|
| 2083 |
3" Tank Extender, 6102/7102 |
|
|
| 5013-L |
Silicone Oil, type 200.20, 1 liter (10 to 230°C) |
|
|
| 9310 |
Carrying Case, 6102 ATA style |
|
|
| 7102 |
| 7102 |
Micro-Bath, -5°C to 125°C (23°F to 257°F) |
|
|
| 2082-P |
Spare Test Lid, 7102 |
|
|
| 2083 |
3" Tank Extender, 6102/7102 |
|
|
| 5010-L |
Silicone Oil, type 200.05, 1 liter (-40 to 300°C) |
|
|
| 9311 |
Carrying Case, 7102 ATA style |
|
|
| 7103 |
| 7103 |
Micro-Bath, -30°C to 125°C (-22°F to 257°F) |
|
|
| 2085 |
Spare Test Lid, 7103 |
|
|
| 5010-L |
Silicone Oil, type 200.05, 1 liter (-40 to 300°C) |
|
|
| 9317 |
Carrying Case, 7103 ATA style |
|
|
|