IR-500
Portable IR Calibrator (Blackbody Target)

Model #IR-500-110

Check/Certify IR thermometers to 932°F (500°C)
Large 2.25 inch (57mm) diameter black body target
Includes 1/8" (3.2mm) diameter well for a precision reference RTD thermometer

Limited Time Price! Was $899

Bookmark and Share

Click and hover to zoom. Click again to view larger image.

Check the performance of infrared guns quickly and easily with our new IR-500. It reaches an upper temperature of 932°F (500°C) in about 40 minutes. You set the display to the desired temperature, then point, shoot and record. It's that easy.

The isothermal 2.25 inch black body target is manufactured to approximate an emissivity of 0.95, which is ideal for most industrial infrared thermometers. For increased accuracy, insert a 1/8" diameter precision RTD into the pre-drilled calibration well and connect to a precision reference meter like our P750.

Each IR-500 includes a NIST-Traceable calibration certificate issued from our own calibration lab. We also include the test data at 122°F and 212°F (50°C and 100°C). An optional 5-point calibration is also available to 752°F (400°C).

The IR-500 provides a stable blackbody source for certifying all types of infrared guns.

Range 122 to 932°F (50 to 500°C)
Accuracy ±1.6°F below 212°F (±0.8 below 100°C)
±3.2°F from 212 to 392°F (±1.6°C from 100 to 200°C)
±5.6°C from 392 to 932°F (±2.8°C from 200 to 500°C)
Resolution 0.1°C/°F
Target Size 2.25" (57mm)
Target Emissivity 0.95
Stability ±0.2°F at 212°F (±0.1°C at 100°C)
±0.8°F at 932°F (±0.4°C at 500°C)
Heating Tme 30 minutes to max
Cooling Time 30 minutes max to 212°F (100°C)
Power 110V AC, 3A or 230V AC (±10%), 1.5A
Dimensions 4.49" x 7.08" x 9.17"
(114 x 180 x 233 mm)
Weight 5.9 lbs (2.682 kg)
Certificate Includes 2-point NIST-Traceable Certificate at 122 & 212°F (50 & 100°C)
Cal Cert

NIST-Traceable IR-500 Calibration

Model: CAL-IR500

  • 5-Points: 122°F, 212°F, 392°F, 572°F, 752°F (50°C, 100°C, 200°C, 300°C, 400°C)
  • Lab calibration using certified sources
5-point calibration performed in ThermoWork's NIST-Traceable calibration lab. Includes calibration report and actual test data.



Our Company