Thermocouple Probes

0.1°C High Accuracy Type T Thermocouple Probes

connectorThe maximum temperature listed for the probes applies to the sensor itself. Plastic handles and thermocouple plugs are rated for short term exposure of 250°F (121°C) and extended exposure of 212°F (100°C).
Thermocouple Probe Quick Selection
Immersion/General Purpose

HT-1, HT-2

Model: HT-1/HT-2

ThermoWorks Image
"Workhorse" probe for liquids, gases, semi-solids. Plastic handle with straight 3 in. stainless shaft. Not good for surface temps. Max Temp. 400°C (750°F). Time constant 0.5 secs. 5 ft. lead. Not isolated.
Range: -58 to 750°F (-50 to 400°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: 0.5 seconds typical in liquid. 2.5 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)
Select probe and add to cart

From:


BT-1

Model: BT-1

ThermoWorks Image
For surface temperatures of solids, also for liquids, gases and semi-solids. Probably our best general purpose probe. Welded stainless-steel shaft, plastic handle. Tip is .028 in. diameter bent at right angle to probe to facilitate accurate surface measurements. 5 ft. lead. Immersible. Max tip temp. 240°C (464°F). Time constant 0.15 secs. Not isolated.
Range: -58 to 464°F (-50 to 240°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: 0.15 seconds typical in liquid. 0.75 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


DPT Series

Model: DPT-12/24/36

ThermoWorks Image
Immersion probes for liquid. #304 stainless steel welded shaft, wooden handle. 5 ft. lead. Max Temp. 400°C (750°F). Time constant 2 secs. Not isolated
Range: -58 to 750°F (-50 to 400°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: 2 seconds typical in liquid. 10 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)
Select probe and add to cart

From:


TFT Series

Model: TFT-12/24/36

ThermoWorks Image
Immersion probes for corrosive liquids. Similar to DPT series but shaft is Teflon® coated. Max Temp. 150°C (300°F). Time constant 3 secs. Isolated. 5 ft. lead.
Range: -58 to 302°F (-50 to 150°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 3 seconds typical in liquid. 15 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)
Select probe and add to cart

From:


PT-6

Model: PT-6

ThermoWorks Image
Sensor Six-Pack. For multi-point measurements. Inexpensive, flexible. Welded sensor beads, tough Kapton insulation, 5 ft. leads. Max Temp. 400°C (750°F). Time constant 0.01 secs. Per package of six. Not isolated.
Range: -58 to 750°F (-50 to 400°C)
Transition (SST): Max temp 750°F (400°C)
Cable: -58 to 750°F (-50 to 400°C)
Time Constant: 0.5 seconds typical in liquid. 2.5 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


GT-1

Model: GT-1

ThermoWorks Image
Air and gas probe. 24 in. #304 stainless steel shaft, wooden handle. 5 ft. lead. For instant readings of temperature gradients in environmental chambers, freezer boxes, etc. Max Temp. 200°C (392°F). Min Temp. –200°C (-328°F). Isolated.
Range: -328 to 392°F (-200 to 200°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: 2 seconds typical in liquid. 10 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


RET-1

Model: RET-1

ThermoWorks Image
Rectal probe for humans, neonatal rabbits, and larger animals. Flexible, vinyl covered, soft tipped. Does not cause discomfort. Max Temp. 90°C (194°F). Time constant 5.0 secs. 5 ft. lead. Isolated.
Range: 14 to 194°F (-10 to 90°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 5 seconds typical in liquid. 25 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


OT-1

Model: OT-1

ThermoWorks Image
For fast reading oral use. Ball-tipped stainless steel shaft, stainless handle. 5 ft. lead. Max Temp. 125°C (257°F). Time constant 0.8 secs. Not isolated.
Range: -58 to 257°F (-50 to 125°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.8 seconds typical in liquid. 4 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


DSP-1

Model: DSP-1

Disposable sterile sheaths for use with OT-1 and RET-1, per 1000.
Range: 32 to 257°F (0 to 125°C)


SST-1

Model: SST-1

ThermoWorks Image
Skin Surface probe, to be taped on. 1/4 in. dia. 10 Kt. gold sensor disc, 5 ft. lead. Max Temp. 90°C (194°F). Time constant 0.15 secs. Not isolated.
Range: 14 to 194°F (-10 to 90°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.15 seconds typical in liquid. 0.75 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


SST-2

Model: SST-2

ThermoWorks Image
Same as SST-1 with plastic handle. Isolated.
Range: 14 to 194°F (-10 to 90°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.15 seconds typical in liquid. 0.75 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


ESO-1

Model: ESO-1

ThermoWorks Image
Esophageal probe. Flexible vinyl covered 1/8 in. dia. Max Temp. 90°C (194°F). Time constant 5.0 secs. 5 ft. lead. Isolated.
Range: 14 to 194°F (-10 to 90°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: 5 seconds typical in liquid. 25 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


RET-2

Model: RET-2

ThermoWorks Image
Rectal probe for rats. For fast intermittent measurements. Smooth ball tip. Stainless steel shaft is 1 inch long and 0.059 inches in diameter. Tip diameter is .125 in. 5 ft. lead. Max Temp. 125°C (257°F). Time constant 0.8 secs. Not isolated.
Range: -58 to 257°F (-50 to 125°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.8 seconds typical in liquid. 4 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


RET-3

Model: RET-3

ThermoWorks Image
Rectal probe for mice, etc. Same as RET-2, with shaft 3/4 inches long and .028 in diameter. Tip diameter is .065 in. Time constant 0.5 secs. Not isolated.
Range: -58 to 257°F (-50 to 125°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.8 seconds typical in liquid. 4 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


Needle Microprobes

Model: MT-x

ThermoWorks Image
Fast-response needle probes for instant readings in tissue, semi-solids, liquids. Also for very small specimens, powders and materials. Needle tip is sealed to ensure only stainless contacts specimen. Max Temp. 200°C (392°F). 5 ft. lead. Smallest microprobes give fastest reading. Short probes are easier to insert and last longer. Item #'s indicate needle gauge and needle length in cm. Sizes Available (ga/cm):MT-29/1, MT-29/2, MT-29/3, MT-29/5, MT-26/2, MT-26/4, MT-26/6, MT-23/3, MT-23/5, MT-23/8.
MT-4 – Similar to MT-29/1 but has blunt tip. Good for instant skin and surface temperatures, liquids. Time constant .025 secs. Not isolated.
All above probes are non-isolated. Isolated versions of 23 gauge and larger needles are available for special order.
Range: -58 to 392°F (-50 to 200°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.025 seconds typical in liquid. 0.125 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)
Select probe and add to cart

From:


IT-23

Model: IT-23

ThermoWorks Image
For ultra fast measurements and for use on micro-size specimens. Max diameter at tip 0.011 in. Tissue implantable with 23 ga. needle (supplied). Rather fragile. Max Temp 150°C (302°F). Time constant 0.005 secs. 3 ft. lead. Teflon™ coated. Isolated.
Range: -58 to 302°F (-50 to 150°C)
Transition (SST): Max temp 302°F (150°C)
Cable: -58 to 302°F (-50 to 150°C)
Time Constant: 0.005 seconds typical in liquid. 0.025 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


IT-1E

Model: IT-1E

ThermoWorks Image
Same as IT-18 but sensor bead exposed. Combines the ultra-fast response of IT-23 with the Teflon™ sheath strength of IT-18. Time constant .005 secs. Isolated.
Range: -58 to 302°F (-50 to 150°C)
Transition (SST): Max temp 302°F (150°C)
Cable: -58 to 302°F (-50 to 150°C)
Time Constant: 0.005 seconds typical in liquid. 0.025 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


MT-D

Model: MT-D

ThermoWorks Image
Fast response surface probe for locating inflammation, arteries, etc. Also for dental use. Stainless steel. 5 ft. lead. Max Temp. 200°C (392°F). Time constant 0.025. Not Isolated.
Range: -58 to 392°F (-50 to 200°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-10 to 105°C)
Time Constant: 0.025 seconds typical in liquid. 0.125 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


ICT-4

Model: ICT-4

ThermoWorks Image
For use on integrated circuits and other micro specimens. 0.013 in. dia. Straight stainless steel shaft; slightly protruding sensor bead to facilitate temperature probing. 5 in. tubular handle fits micro manipulators. Max. intermittent temp. 200°C (392°F), continuous use 100°C (212°F), Time constant 0.025 secs. 5 ft. lead. Not isolated.
Range (intermittent): -58 to 392°F (-50 to 200°C)
Range (continuous): -32 to 212°F (0 to 100°C)
Transition (SST): Max temp 221°F (105°C)
Cable: 14 to 221°F (-14 to 105°C)
Time Constant: .025 seconds typical in liquid. 0.125 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)


Insect Probe

Model: MT-29/1B

ThermoWorks Image
Similar to MT-29/1 but sensor is swaged into tip for max heat transfer. Designed for high accuracy on extremely small specimens such as insects, seeds, etc. Max. insertion depth 1/8 in. Time constant 0.015 secs. 5 ft. lead. Max Temp. 150°C (302°F). Other sizes made to special order. Not isolated.
Range: -58 to 302°F (-50 to 150°C)
Transition (SST): Max temp 302°F (150°C)
Cable: -58 to 302°F (-50 to 150°C)
Time Constant: 0.015 seconds typical in liquid. 0.075 seconds full reading.
Accuracy: ±0.2°F (±0.1°C) from 32 to 122°F (0 to 50°C)

Icon Legend
Icon Category Icon 1
Icon Category Icon 2
Icon Category Icon 3
Icon Category Icon 4
Icon Category Icon 5
Icon Category Icon 6

Our Company